SIGNAL RECOVERY
productsblue.jpg (1564 bytes) orderonlineblue.jpg (1564 bytes) searchblue.jpg (1470 bytes) supportblue.jpg (1492 bytes) newsblue.jpg (1518 bytes) contactusblue.jpg (1558 bytes)    
...see what you've been missing
blue2pxdot.jpg (633 bytes)
Digital Signal Averagers Index
 
 
 

Product Selection Guide:-
Digital Signal Averagers

 

Feature Model
  FASTFLIGHT-P FASTFLIGHT-2
Max. Spectra/s in Chromatograph/TOF-MS Mode 10 spectra/s 100 spectra/s

Sampling Intervals

Interpolated

-

0.25 ns

Interleaved

0.5, 1 ns

0.25 ns

Real time

2 ns

0.5, 1, 2 ns


Sampling Clock Temperature Sensitivity

within ±10 ppm per
degree C

within ± 2.5 ppm,
0 to 50 C


Max. Number of Time Bins

262,000

1,500,000


Min./Max. Time Spans

0.25 ns

-

10 / 375 µs

0.5 ns

0.256 / 131 µs

10 / 750 µs

1.0 ns

0.512 / 262 µs

10 / 1500 µs

2.0 ns

1.024 / 524 µs

10 / 3000 µs


Data Acquisition Delay

0 to 1048 µs

0 to 1048 µs


Dead Time

End-of-Scan

0.544 to 8.544 µs

0.8 µs

End-of-Spectrum

0.544 to 8.544 µs

0.8 µs


Analog Input Range (full scale)

0 to -1 V

 

0 to -0.5 V


Analog DC Offset Range

-400 mV to +400 mV

-250 mV to +250 mV


Sampling ADC

8 bits

8 bits


Amplitude Resolution with Precision Enhancer

12 bits

12 bits


Equivalent Input Noise

Uncorrelated

<2 mV rms

<2 mV rms

Correlated

<0.2 mV rms

<0.02 mV rms


Automatic Correlated Noise Subtraction

No

Yes


ADC Non-linearity

Differential

within ±0.15 LSB

within ±0.10 LSB

Integral

within ± 0.4% of full scale

within ± 0.4% of full scale


Triggering Options

Trigger Output

Trigger Input or Output


Duet Operation

No

Yes


Data Acquisition Modes

a) TOF only

a) TOF only

b) Chromatograph/TOF-MS

b) Chromatograph/TOF-MS


Data Compression Modes

a) Uncompressed

a) Lossless

b) Peak-preserving & background-decimating

b) Peak-preserving & background-decimating

c) Peak area & centroid only.


Rapid Protocol Switching Time

<1 ms

10 µs


Package Size

427 mm W x 396 mm D x 81 mm H

330 mm W x 340 mm D x 74 mm H


Interface

Cable and PCI card

USB-2


CE Emissions Level

A

B


Software Operating System

Windows 2000/XP

Windows XP


Programmer's Toolkit

Included

Included

 


SIGNAL RECOVERY is part of Advanced Measurement Technology, Inc, a division of AMETEK, Inc.
 It includes the businesses formerly trading as EG&G Instruments (Signal Recovery),
EG&G Signal Recovery and PerkinElmer Instruments (Signal Recovery)

Copyright© 2010 AMETEK, Inc.  Legal Notices

801 South Illinois Avenue, Oak Ridge, TN 37831-2011, U.S.A. (865) 482-4411 All Rights Reserved.

Ametek logois a trademark of AMETEK, Inc.