Specifications
Instruction Manual
 
 
 

Models 9305-P
Fast Preamplifier


  • Fast preamplifier for use with photomultipliers, electron multipliers, etc., in photon or ion counting applications
  • Output of ±5 V into 50 W
  • Input protection
  • <3 ns rise and fall times
  • DC coupled, with excellent DC offset and gain stability
  • Voltage gain adjustable from x5 to x10
  • <5 ns overload recovery time
  • Compact and lightweight

General
The Model 9305 Fast Preamplifier contains a direct-coupled wideband hybridized amplifier suitable for use with photomultipliers, electron multipliers, and other detectors used in photon counting, ion counting, or fast-timing applications.

In addition to the fast rise time (<3 ns), the Model 9305 has a variable voltage gain of x5 – x10 and can drive ±5 V into a 50 W load. The 9305 also features excellent DC offset and gain stability along with low noise and <5 ns overload recovery time. Overload input protection is provided also.

Hybrid circuit technology gives the Model 9305 high performance and reliability. For operator convenience independent controls for bandwidth (BDW), output DC offset (DC), and gain (GAIN) are included.

Because of its compact size and light weight, the Model 9305 is ideal for mounting directly on or close to a detector. Consequently, low-level signals which would otherwise be susceptible to distortion by noise or interference are amplified to a suitable level for cable connection to the main amplifier.

The instrument is powered via an integral 3 m (10') long power cable fitted with a 9-pin D type connector that is compatible with ORTEC NIM-standard modules.

Input  
Connector BNC
Mode Single-ended
Coupling DC
Impedance 50 W
Noise £ 25 µV rms equivalent input noise
Gain x5 to x10 (screwdriver adjustable), non-inverting
Gain Accuracy ±2%
Gain Stability < ±0.05%/°C.
Output
Connector BNC
Impedance < 1 W
Voltage Will deliver ±5.0 V into a 50 W load
Coupling DC
DC Stability £ ±150 µV/°C referred to the output
Risetime £ 3 ns
Bandwidth > 120 MHz
Linear Range ±5 V
Nonlinearity < ±1% for output voltages up to ±5 V
Overload Recovery 5 ns typical for x10 overload
Controls
BDW 20-turn front-panel screwdriver adjustable potentiometer sets the output bandwidth and overshoot
DC 20-turn front-panel screwdriver adjustable potentiometer sets the output DC offset
GAIN 20-turn front-panel screwdriver adjustable potentiometer sets the voltage gain in the range x5 to x10
Power ±12 V regulated DC @ 67 mA max, via 9-pin D type connector. Supplied with integral 3 m (10') power cable.
Dimensions (excluding connectors) 38 mm wide x 89 mm deep x 61 mm high
(1.5" wide x 3.5" deep x 2.4" high)
Weight Net 320 g (0.75 lb)
Shipping 1.25 kg (2.75 lb)


SIGNAL RECOVERY is part of Advanced Measurement Technology, Inc, a division of AMETEK, Inc.
 It includes the businesses formerly trading as EG&G Instruments (Signal Recovery),
EG&G Signal Recovery and PerkinElmer Instruments (Signal Recovery)

Copyright© 2008 AMETEK, Inc.  Legal Notices

801 South Illinois Avenue, Oak Ridge, TN 37831-2011, U.S.A. (865) 482-4411 All Rights Reserved.

Ametek logois a trademark of AMETEK, Inc.