SIGNAL RECOVERY
productsblue.jpg (1564 bytes) orderonlineblue.jpg (1564 bytes) searchblue.jpg (1470 bytes) supportblue.jpg (1492 bytes) newsblue.jpg (1518 bytes) contactusblue.jpg (1558 bytes)    
...see what you've been missing
blue2pxdot.jpg (633 bytes)
Specifications
Instruction Manual
 
 
 

Model 4002D-P
160 W Power Supply for NIM Bins

4006SR34Medium.jpg (13145 bytes)

  • Regulated dc power of ±6 V @ 10 A, ±12 V @ 3 A, and ±24 V @ 1.5 A; 115 V AC @ 0.5 A
  • 160 W of dc output up to 50°C
  • Short-circuit and overload protected
  • Over-voltage protection to avoid damage to +5 V and –5.2 V integrated circuits
  • Operates from 100, 115, 200, or 22 V AC at 47 to 63 Hz

General
T
he Model 4002D-P Power Supply is designed to supply DC power to a NIM bin when the application requires ±6 V, ±12 V, and ±24 V power. The 4002D-P Power Supply can be purchased separately for use with existing NIM bins, or it can be ordered attached to either a 4001A-P or 4001C-P NIM Bin. Mounting hardware is supplied to make the Model 4002D-P compatible with all ORTEC NIM bins. The Model 4001C-P NIM Bin is recommended for use with Model 4002D-P because Model 4001C-P distributes the power with copper bus bars to minimize the voltage drop at each module's power plug. The 4002D-P Power Supply is designed to exceed recommended power supply specifications for Type V-H supplies as defined in DOE/ER-0457T.

Regulated DC power supplied to the attached bin by Model 4002D-P is conservatively rated at +6 V @ 10 A, – 6 V @ 10 A, +12 V @ 3 A, –12 V @ 3 A, +24 V @ 1.5 A, and –24 V @1.5 A. These maximum output currents can be delivered in any combination, provided the total output DC power does not exceed 160 W at ambient temperatures up to 50°C. In addition, 115 V AC is available up to 0.5 A.

Protection against overload is provided in several ways. When the heat sink temperature exceeds 95°C, the red warning indicator is illuminated on the attached bin control panel. When the heat sink temperature exceeds 110°C, the power supply is automatically shut down, causing both the power and temperature indicator lights to turn off. Recovery from thermal overload is automatic when the thermal load is reduced. Output currents from the DC supplies are internally limited to 120% of their rated values by foldback circuits. This provides overload and short-circuit protection. On the +6 V and –6 V DC supplies, crowbar circuits limit the output voltage to 7.5 V to protect integrated circuits. Fuses protect the AC inputs to the power supply.

An external slide switch allows selection of either 115 or 220 V AC as the power input. By changing pins on an internal connector, this selection can be altered to 100 and 200 V AC. A standard IEC line input connector allows power cords and plugs that meet local electrical standards to be used. Control of the primary power is provided by the On/Off switch on the NIM bin control panel.

Connection of power and control lines to the NIM bin is provided by the standard interface connector specified in DOE/ER-0457T. Mechanical mounting of the power supply to the bin is with brackets utilizing the standard bolt pattern specified in DOE/ER-0457T.

Power Supply  
Input 103–129 or 200–258 V AC, 47–63 Hz. An external slide switch selects nominal input voltages of 115 or 220 V AC. Changing pins on an internal connector allows operation at 88–110 V or 191–239 V AC, 47–63 Hz, with the external slide switch selecting nominal voltages of 100 or 200 V AC. Input current at 115 V AC is nominally 4 A for a 160 W dc output simultaneous with a 0.5 A, 115 V AC output. Dual fuse input uses 8-A SB U.S.A. standard fuses for 100 or 115 V AC, 60 Hz and 5 A SB metric fuses for 200 and 220 V AC, 50 Hz operation
DC Outputs Maximum rated currents for each voltage supplied to the 6 module power connectors and 2 preamplifier power connectors are:
DC Voltage Maximum Current DC Voltage Maximum Current
+24 V  1.5 A –24 V  1.5 A
+12 V 3 A –12 V 3 A
+6 V 10 A  –6 V 10 A

Maximum total dc output power: 120 W at 23°C, 160 W up to 50°C ambient air temperature. Derate 3%/°C for 50 to 60°C

115 V AC Output Unregulated voltage. Maximum current limited only by the input fuses when operated in the 100- or 115-V AC settings. Limited to 0.5 A on the 200- and 220-V AC settings when the DC load is 160 W. Output voltage is nominally 115 V AC in the 115-V and 220-V input modes. Output voltage is nominally 100 V AC in the 100 V and 200 V input modes.
Regulation <±0.1% (typically ±0.05%) for ±12 V and ±24 V, and <±0.2% (typically ±0.1%) for ±6 V over the combined range of zero to full load with the specified input voltage range for measurements made within a 1-minute period. Regulation <±0.3% for ±12 V and ±24 V, and <±0.6% for ±6 V over any 24-hour period at constant ambient temperature for the same load and input ranges after a 60-minute warm-up.
Stability < DC output voltages change <±0.5% (after a 60-minute warmup) over a 6-month period at constant load, line voltage, and ambient temperature
Output Impedance <0.3 W at any frequency up to 100 kHz for the DC outputs
Temperature Coefficient <0.02%/°C from 0 to 60°C
Noise & Ripple <3 mV peak-to-peak, as observed on 50 MHz bandwidth oscilloscope
Voltage Adjustment ±2% minimum range. Resettability <±0.05% of the supply voltage
Recovery Time <100 µs to return to within ±0.1% of the rated voltage for all dc outputs for any input voltage change within the rated range or for a change of load current from 10% to 100% of full load.
Protection Both input power lines include fuses. The power supply is automatically turned off by an internal switch if the temperature of the heat sinks exceeds 110°C. Recovery is automatic when the temperature decreases to a safe value. Provision is made for activating a temperature warning light on the NIM bin control panel to advise that the temperature limit is being approached. This warning occurs at and above a heat sink temperature of 95°C. All dc outputs include a current foldback circuit to limit the output current to nominally 120% of the rated value. This feature provides short-circuit and overload protection. Recovery is automatic after removal of the overload condition. Over-voltage protection for the ± 6 V outputs prevents these outputs from exceeding ±7.5 V, respectively, to protect the integrated circuits that are commonly powered by these supply voltages
Dimensions 89 mm high x 432 mm wide x 269 mm deep
(3.5" high x 17.0" wide x 10.6" deep)
Weight Net 11.3 kg (25 lb)
Shipping 15.9 kg (35 lb)


SIGNAL RECOVERY is part of Advanced Measurement Technology, Inc, a division of AMETEK, Inc.
 It includes the businesses formerly trading as EG&G Instruments (Signal Recovery),
EG&G Signal Recovery and PerkinElmer Instruments (Signal Recovery)

Copyright© 2010 AMETEK, Inc.  Legal Notices

801 South Illinois Avenue, Oak Ridge, TN 37831-2011, U.S.A. (865) 482-4411 All Rights Reserved.

Ametek logois a trademark of AMETEK, Inc.