Reply Next Previous Up

Re: 7265 RS232 Communication

From: Signal Recovery
email: info@signalrecovery.com
Model: 7265
Date: 12 Aug 2005

Comments

As advised, I suspect you need a null modem cable. The 7265 RS232 port is configured as in the same way as a PC so you need a null modem to connect it. A lot of other RS232 instruments are set as a terminal, and so work with a straight-through cable. Our part number for a suitable cable is the CO1003. The interconnection diagram is on our website at http://www.signalrecovery.com/faqs/00000011.htm


Last changed: April 01, 2008