SIGNAL RECOVERY
HOME PRODUCTS ORDER
ONLINE
SEARCH SUPPORT NEWS CONTACT
US
 
...see what you've been missing
  


Model 5188B
Optical Input Preamplifier

Model 5188B Optical Input Preamplifier

General
The model 5188B preamplifier has a silicon photodiode input and is designed to be used as a light to voltage converter where the signal to be amplified is derived from an optical source in the visible region of the electromagnetic spectrum. Sensitivity is switch-selectable, with a choice of five settings, which enables the amplifier to detect fractions of a picowatt, without noise degradation, on its most sensitive range, while being able to accept optical power of up to 18 mW on its least sensitive range without overload.

The instrument has bandwidths from 25 kHz on its most sensitive range to 1 MHz on its least sensitive range. Input noise ranges from 180 fW/ÖHz to 25 pW/ÖHz, depending on the sensitivity setting selected. An additional low-noise range, denoted 10-8 W/V LOW NOISE, is also provided, this setting having a reduced bandwidth of 10 kHz.

Input signal connection is made via an FC/PC optical connector on the front panel of the unit. AC and DC components of the input signal are processed such that independent outputs for each are made available via BNC connectors, also on the front panel. This will allow users to measure modulated and CW components of an input signal separately and perform, for example, ratios between them.

The model 5188 can be powered from its own internally housed (alkaline) batteries, an external low voltage supply (±15 V or ±18 V) or from the model PS0108 remote line power supply (optional extra). This preamplifier can also be powered from most of our range of lock-in amplifiers and from the model 7310 noise rejecting voltmeter.

Input
Connector FC/PC
Device Type
    5188B Si
Operating Wavelength
    5188B 400 nm to 1000 nm
Sensitivity Switch selectable (5 settings)
    AC Output 10-5, 10-6, 10-7, 10-8, 10-8 low noise W/V
    DC Output 10-3, 10-4, 10-6, 10-7, 10-8 W/V
Calibration Factor
    5188B Sensitivity setting x 2.2 at 830 nm
Accuracy ±2%
Stability ±300 ppm/degree C
Frequency Response
    10-5 W/V setting 0.5 Hz to 1 MHz
    10-6 W/V setting 0.5 Hz to 800 kHz
    10-7 W/V setting 0.5 Hz to 200 kHz
    10-8 W/V setting 0.5 Hz to 25 kHz
    10-8 W/V low noise
    setting
0.5 Hz to 10 kHz
Max Optical Power at input
    10-5 W/V setting 518 mW
    10-6 W/V setting 1.8 mW
    10-7 W/V setting 18 µW
    10-8 W/V setting 1.8 µW
    10-8 W/V low noise
    setting
180 nW
Output
AC Output
    Impedance 450 W
    Max voltage swing > 10 V pk-pk
    Slew rate > 22 V/µs
Noise, rms. referred to input
    10-5 W/V setting 25 pWÖHz
    10-6 W/V setting 15 pWÖHz
    10-7 W/V setting 350 fWÖHz
    10-8 W/V setting 180 fWÖHz
    10-8 W/V low noise
    setting
150 fWÖHz
DC Output
    Impedance 10 kW
    Max voltage swing > ±9 V
Output due to photodiode dark current, max
    10-3 W/V setting 10 µV
    10-4 W/V setting 100 µV
    10-6 W/V setting 10 mV
    10-7 W/V setting 5100 mV
    10-8 W/V 51 V
Power Internal Four 9 V alkaline batteries provide approximately 15 hours of use
External a) ±15 V or  ±18 V DC @ 25 mA
b) 110 V AC or 240 V AC via optional external model PS0108 power supply
Dimensions (excluding connectors) 8.25" wide x 11" deep x 3.5" high
(210 mm wide x 279 mm deep x 89 mm high)
Weight 5.3 lbs. (2.4 kg) excluding power supply

 


SIGNAL RECOVERY is part of Advanced Measurement Technology, Inc, a division of AMETEK, Inc.
 It includes the businesses formerly trading as EG&G Instruments (Signal Recovery),
EG&G Signal Recovery and PerkinElmer Instruments (Signal Recovery)

Copyright© 2008 AMETEK, Inc.  Legal Notices

801 South Illinois Avenue, Oak Ridge, TN 37831-2011, U.S.A. (865) 482-4411 All Rights Reserved.

Ametek logois a trademark of AMETEK, Inc.