Model 5182
Current Sensitive Preamplifier

Model 5182 Current Preamplifier

General
The model 5182 is a current-to-voltage preamplifier of low noise and low input impedance designed to amplify the extremely low  low currents encountered in such areas as photometry and semiconductor research. It has four standard conversion factors but in addition includes a special low-noise mode on the highest gain position for even better low current measurement capability. The unit features two outputs, allowing both the AC and DC components of the input signal to be independently monitored, so that, for example, in a PMT application the bias current can be measured separately from the signal current.

It can be powered from its own internally housed (alkaline) batteries, an external low voltage supply (±15 V or ±18 V) or from the model PS0108 remote line power supply (optional extra). This preamplifier can also be powered from most of our range of lock-in amplifiers and from the model 7310 noise rejecting voltmeter

The model 5182 is ideally suited to amplifying signals from current sources such as electron multipliers, ion collectors, photomultipliers and photodiodes.

Frequency Response (typical)
Model 5182 Frequency Response Plot
Input Impedance (typical)
Model 5182 Input Impedance Plot


Peak Input Current and Noise Current (typical)

Gain
A/V
Max DC Input Current Noise Current at 1 kHz
10-5 9 mA 10 pA/root Hz
10-6 900 µA 5 pA/root Hz
10-7 9 µA 135 fA/root Hz
10-8 900 nA 45 fA/root Hz
10-8, low noise 90 nA 15 fA/root Hz

Input
Configuration BNC connector, virtual ground.
Coupling DC
Sensitivity Switch selectable (5 settings)
    AC Output 10-5, 10-6, 10-7, 10-8, 10-8 low noise A/V
    DC Output 10-3, 10-4, 10-6, 10-7, 10-8 A/V
    Accuracy ±2%
    Stability ±300 ppm/degree C
Impedance see plot above
Frequency Response lower limit 0.5 Hz; upper limit depends on sensitivity setting, see plot above
Max DC current at input see table above
Max input without damage ±15 V DC or 10 V rms. AC @ 50 Hz
Noise see table above
Output
AC Output
    Impedance 450 Ω
    Max voltage swing > 10 V pk-pk
    Slew rate > 22 V/µs
DC Output
    Impedance 10 kΩ
    Max voltage swing > ±9 V
Polarity Current flowing into the input produces a positive output voltage
Power Internal Four 9 V alkaline batteries provide approximately 15 hours of use
External a) ±15 V or  ±18 V DC @ 25 mA
b) 110 V AC or 240 V AC via optional external model PS0108 power supply
Dimensions (excluding connectors) 8.25" wide x 11" deep x 3.5" high
(210 mm wide x 279 mm deep x 89 mm high)
Weight 5.3 lbs. (2.4 kg) excluding power supply

 


SIGNAL RECOVERY is part of Advanced Measurement Technology, Inc, a division of AMETEK, Inc.
 It includes the businesses formerly trading as EG&G Instruments (Signal Recovery),
EG&G Signal Recovery and PerkinElmer Instruments (Signal Recovery)

Copyright© 2010 AMETEK, Inc.  Legal Notices

801 South Illinois Avenue, Oak Ridge, TN 37831-2011, U.S.A. (865) 482-4411 All Rights Reserved.

Ametek logois a trademark of AMETEK, Inc.